AM-5100 SO2 analyzer adopts ultraviolet fluorescence technique and combines advanced micro processing technology, which could provide accurate and reliable SO2 detection at low level ppb~ppm.This analyzer can be applied to atmospheric monitoring and dilution-based CEMS system.
Features Of AM-5100 SO2 Analyzer
Measurement range 0~500ppb to 0~20ppm, user selectable;
High efficiency hydrocarbon remover and unique optical design, strong anti-interference ability;
Dark current compensation at fixed time, high stability;
Self checking function of light intensity attenuation;
Adaptive signal filtering technology to optimize response time;
Large capacity memory, automatic storage of historical data.
Application Of AM-5100 SO2 Analyzer
Atmospheric monitoring
Specification Of AM-5100 SO2 Analyzer
Range:0~500ppb, wide range 0~20,000ppb (optional)
Unit:ppb, ppm, ug/m3, mg/m3
Zero Point Noise:<0.2ppb (RMS)
Low Detection Limit:0.4ppb
Zero Drift:<1ppb/24h
Span Drift:<5ppb/24h
Response Time:<120s (T90)
Linearity:<1%/F.S.
Repeatability:<1%
Sampling Flow Range:800SCCM±10%
Ambient Temp.:5℃~40℃
Dimension:178mm×430mm×635mm
Weight:<15kg, external pump 5kg
Power:200V~240V, 50/60Hz, 400W
Analog Output:10V, 5V, 1V, 0.1V (optional)