HM-100 High-temperature Hygronom

(2020年02月17日)

http://www.zetian-group.com/products/hm-100-high-temperature-hygronom/

Overview
Hygronom is also called moisture meter, gas moisture meter or dew-point meter, which is used to measure the humidity of flue gas emitted by pollution source. According to measured humidity, it converts wet gas concentration into dry gas concentration. In addition, it also can be used in industrial online humidity monitoring.

HM-100 flue gas hygronom is equipped with an internal high-precision capacitance digital temperature & humidity sensor. The installation method adopts the in-situ immersion installation. It is a self-developed product, of which, the core component is from Switzerland. Put the sensor inside stack to get in touch with flue gas. The probe needs preheating to prevent gas from condensation, which will affect the measurement of moisture in the flue gas. After gas contacts with the measurement chip, the chip will change capacitance to measure flue gas concentration.

Features Of HM-100 High-temperature Hygronom

• The product is convenient for installation, disassembly, and cleaning with small and compact structure
• Direct and simple key operation on site 
• The probe is equipped with heat tracing function to prevent condensate water from damaging humicap and thermal resistance 
• Imported core components with high-performance index and reliability

Technical Parameter Of HM-100 High-temperature Hygronom

Measurement RangeH2O: 0~40.00%(volume percent)(optional)
Humidity sensorResistance-capacitance
Temp sensorPt100 platinum resistor
Accuracy0 ~ 40.00% H2O: ± 2%
Repeatability0 ~ 40.00% H2O: ± 1%
Stability0 ~ 40.00% H2O: ± 1% FS / 7d
Response timeT90 < 6s
Sensor life>2 years (under normal use)
Transmitter life>5 years (under normal use)
Power supply220VAC
Instrument weightAbout 10kg
Power consumptionMaximum about 35VA (heating)
Ambient temperatureTransmitter: -10°C~+55°C
Probe: 0~180°C (except probe cable)
Ambient humidity<95%RH
Sample gas flow(2 ~ 6) L / min
Analog output (2 sets)Temperature and humidity, 4-20mA
CommunicationRs232 (standard)

Typical Application Of HM-100 High-temperature Hygronom

1.High-temperature humidity measurement in exhaust emission (CEMS) of fire coal and oil burning boiler 
2.Humidity measurement in food processing
3.Humidity measurement in wood, building material, and papermaking industry
4.Humidity measurement in fields of chemical industry and fiber
5.Humidity measurement under high temperature of open width dyeing in dyeing & printing industry

Hangzhou Zetian Technology is committed to the integrity, strength and quality of our water quality testing products. We're internationally recognized. Welcome you to visit us.

hm-100-humidity-analyzer.png

コメント